Model Number 4FC12
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/21/2016
Event Type  malfunction  
Event Description
It was reported that during a cryoablation procedure, air bubbles were observed when aspirating the sheath.It was noted that the balloon catheter was introduced into the sheath and was forwarded into the atrium.Every time aspiration was attempted, air bubbles were visible.The sheath was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the sheath was returned and visually inspected and functionally tested.Clinical data files were received and reviewed; the files showed at least 18 injections were performed with without any issue on the date of the event.Visual inspection of the sheath showed the device was intact with no apparent issues.Air aspiration was reproduced when a test catheter was introduced through the sheath.Dissection showed the hemostatic valve was leaking.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA   H9H 5H3
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6136796
MDR Text Key61193888
Report Number3002648230-2016-00507
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/27/2018
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number09404
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-