Catalog Number 595000-001
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/24/2016
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the freedom driver exhibited a fault alarm, it continued to perform its life-sustaining functions.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited a fault alarm while supporting a patient.The customer also reported that the patient was switched to a back-up freedom driver without any reported adverse patient impact.
 
Manufacturer Narrative
The freedom driver was returned to syncardia for evaluation.The driver passed all pressure test requirements, which included cardiac output, rap (right atrial pressure), aop (aortic pressure), pap (pulmonary arterial pressure) and lap (left atrial pressure) performance metrics associated with nominal normotensive and hypertensive settings.Review of the electronic data revealed a fault code of "bottom pressure too low" which is indicative of a potentially failing u22 pressure sensor on the main pcba (printed circuit board assembly) and is consistent with the alarm that the customer experienced.Further investigation of the main pcba confirmed that the root cause for the reported fault alarm was a degraded pressure sensor (u22) on the main pcba.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited a fault alarm while supporting a patient.The customer also reported that the patient was switched to a back-up freedom driver without any reported adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6141112
MDR Text Key61601845
Report Number3003761017-2016-00380
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 11/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/24/2016
Initial Date FDA Received12/02/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/04/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/02/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age34 YR
-
-