Model Number 595000-001
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/24/2016
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the freedom driver exhibited a fault alarm, it continued to perform its life-sustaining functions.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a (b)(6) hospital, reported that the freedom driver exhibited a fault alarm while supporting a patient.The customer also reported that the patient was switched to a back-up freedom driver without any reported adverse patient impact.
 
Manufacturer Narrative
The freedom driver was returned to syncardia for evaluation.The customer-reported fault alarm could not be duplicated through investigational testing; therefore, the root cause could not be conclusively determined.The investigation determined there was no evidence of a device malfunction and the driver performed as intended.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited a fault alarm while supporting a patient.The customer also reported that the patient was switched to a back-up freedom driver without any reported adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6141117
MDR Text Key61606916
Report Number3003761017-2016-00379
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 11/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/24/2016
Initial Date FDA Received12/02/2016
Supplement Dates Manufacturer Received09/11/2017
Supplement Dates FDA Received09/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age34 YR
-
-