Catalog Number ASKU
Device Problems Air Leak (1008); Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
No sample has been returned for evaluation; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted.Due to an observed increased trend for this event, an internal investigation was initiated to determine if corrective and preventative actions were necessary.A root cause for the increased complaint rate was found to be related to a manufacturing post assembly inspection practice.This complaint does not identify a reported lot and no sample was returned for evaluation so it cannot be determined if the complaint can be attributed to the post assembly inspection.The post assembly inspection has been discontinued and an effectiveness check has been established and will be monitored periodically.(b)(4).
 
Event Description
An ophthalmic surgeon reported that at least one of the three valved trocars leaked upon insertion in over fifty percent of the vitrectomy procedures for retinal detachments.In instances where the patient was already vitrectomized and there was gas or liquid present in the eye, one might get leakage immediately."often the liquid squirts up toward the biomicroscope lens (biom) lens in one squirt".In addition, "when one switches to air and must e.G.Laser treat in air then it bubbles with air around the laser cable and up under the biom lens which means that one must continuously stop and clean up under the lens".There is no known harm to the patients.Additional information was requested; however, none has been received to date.
 
Manufacturer Narrative
(b)(4).
 
Event Description
Additional information was received from the surgeon.The surgeon indicated in some cases, the leakages were observed as soon as the procedures were started, while in other procedures, leakage occurred after a period of time.In most of the procedures, the leakage was from multiple trocars.The trocars were replaced in a majority of the cases and all the procedure were completed with no harm to the patients.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6144130
MDR Text Key61706222
Report Number1644019-2016-01495
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-