• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP FT4; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP FT4; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number N/A
Device Problems Calibration Problem (2890); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/27/2016
Event Type  malfunction  
Manufacturer Narrative
The cause for the discordant ft4 results is unknown.Siemens healthcare diagnostics is investigating.The ifu states in the interpretation of results section: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings.".
 
Event Description
A false low advia centaur xp ft4 result was obtained for a patient sample.The patient sample was tested at another site for comparison and the result was higher.The tsh3-ultra result was normal.The result from the other site was more aligned with the clinical picture.Patient treatment was not prescribed or altered.There was no report of adverse health consequences due to the discordant ft4 result.
 
Manufacturer Narrative
Siemens filed the initial mdr 1219913-2016-00225 on december 5, 2016.On 01/05/2017 additional information: siemens has performed an internal investigation and has confirmed a negative bias for the advia centaur ft4 when used with calibrator a kit lots ending in 90 on the advia centaur, advia centaur xp and advia centaur xpt systems.In addition, siemens healthcare diagnostics confirmed the potential for calibration failures due to above limit calibrator rlu %cvs when using calibrator a kit lots ending in 90 with the ft4 assay.Siemens issued ufsn cc 17-04.A.Ous (january 2017) and umdr cc 17-04.A.Us (january 2017) informing the customer of ft4 assay negative bias observed with calibrator a kit lots ending in 90.Instructions on actions to be taken are provided in the customer communication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XP FT4
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
511 benedict avenue
tarrytown NY 10591 5097
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
333 coney street
e. walpole MA 02032
Manufacturer Contact
eiman sulieman
333 coney street
e. walpole, MA 02032
5086604603
MDR Report Key6145579
MDR Text Key61452710
Report Number1219913-2016-00225
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132249
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Medical Technologist
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 01/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/25/2017
Device Model NumberN/A
Device Catalogue Number10282219
Device Lot Number49905073
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/05/2016
Initial Date FDA Received12/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/26/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/25/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1219913-01/10/2017-002-R
Patient Sequence Number1
-
-