• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC. LAP-BAND AP SMALL SYSTEM WITH RAPIDPORT EZ; ADJUSTABLE GASTRIC BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC. LAP-BAND AP SMALL SYSTEM WITH RAPIDPORT EZ; ADJUSTABLE GASTRIC BAND Back to Search Results
Model Number C-20360
Device Problem Fluid/Blood Leak (1250)
Patient Problem Failure of Implant (1924)
Event Date 11/07/2016
Event Type  malfunction  
Manufacturer Narrative
Taper ii: the reporter of the event was asked to return the product for analysis.To date, apollo has not received the device.Based on the serial number provided, the connector type is assumed to be a taper ii.If returned, visual examination may confirm or determine another taper type associated with this event.Device labeling addresses the reported event of leak as follows: precautions: care must be taken during band adjustment to avoid puncturing the tubing that connects the access port and band, as this will cause leakage and deflation of the inflatable section.Failure to create a stable, smooth path for the access port tubing, without sharp turns or bends, can result in tubing breaks and leakage.In order to avoid incorrect placement, the port should be placed lateral to the trocar opening.A pocket must be created for the port so that it is placed far enough from the trocar path to avoid abrupt kinking of the tubing.The tubing path should point in the direction of the access port connector so that the tubing will form a straight line with a gentle arching transition into the abdomen.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Deflation of the band may occur due to leakage from the band, the port or the connecting tubing.Warnings: patients should be advised that the lap-band ap® system is a long-term implant.Explant and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
 
Event Description
Reported as: a patient with the lap-band system was reported to have a "leaking port." port was removed and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND AP SMALL SYSTEM WITH RAPIDPORT EZ
Type of Device
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s capitol of texas hwy
bldg 1, ste 300
austin TX 78746
Manufacturer (Section G)
ALLERGAN
global park free zone
900 global park
la aurora de heredia, costa rica
CS  
Manufacturer Contact
laura leboeuf
1120 s capitol of texas hwy
bldg 1, ste 300
austin, TX 78746
5122795141
MDR Report Key6147796
MDR Text Key61559856
Report Number3006722112-2016-00374
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date09/17/2016
Device Model NumberC-20360
Device Catalogue NumberC-20360
Device Lot Number2655572
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/17/2014
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NAPROXEN; VITAMINS
Patient Age25 YR
Patient Weight118
-
-