This is filed to report the difficulty removing the clip delivery system (cds) through the steerable guide catheter (sgc), which has the potential to cause or contribute to patient injury.It was reported that on (b)(6) 2016, the patient, with mixed etiology mitral regurgitation (mr), underwent a mitraclip procedure.During the mitraclip procedure, the clip was deployed.The gripper lines were noted to be difficult to remove, as if there was a lot of tension.The lines were pulled slowly and the "m" knob was released.The device was pulled medially to align the device with the clip in an effort to relieve some of the tension.The gripper lines were removed and clip deployment was completed.The cds was retracted into the sgc and resistance was noted.Force was applied, but the cds could not be removed.Fluoroscopy noted that the cds sleeve had a small "waist", as if it were being crunched.The cds was advanced 1-2 cm back into the left atrium, then pulled back aggressively.The cds was able to be removed without patient injury.Upon removal, an indentation was seen at the end of the steerable sleeve and no damage was noted with the sgc.A second sgc was used to deploy a second planned clip, post procedure, the mixed etiology mitral regurgitation was reduced from grade 4 to grade 1.No additional information was provided.
|
(b)(4).Evaluation summary: the device was returned for analysis and although the reported resistance removing the clip delivery system (cds) from the steerable guide catheter (sgc) could not be confirmed, abbott vascular (av) confirmed resistance at the distal end of the sgc, when inserting the cds into the sgc.A search of the complaint handling database was performed and there were no other complaints identified from this lot related to difficulty removing the cds from the sgc.Further assessment was performed per site operating procedures and it was determined that there is no indication that the design or manufacture of the device contributed to the reported event.Av will continue to trend the performance of these devices.
|
(b)(4).Evaluation summary: the device was returned for analysis and although the reported resistance removing the clip delivery system (cds) from the steerable guide catheter (sgc) could not be confirmed, abbott vascular (av) confirmed resistance at the distal end of the sgc, when inserting the cds into the sgc.A search of the complaint handling database was performed and there were no other complaints identified from this lot related to difficulty removing the cds from the sgc.Further assessment per site operating procedures was performed and av identified a potential product deficiency related to the manufacture of the device.There is no evidence to indicate that a wider population of product has potentially been impacted.Av will continue to trend the performance of these devices.
|