It was reported that the patient passed away after being found unresponsive.The patient¿s wife called ems who initiated cpr, but later the patient was pronounced dead.The device was interrogated and upon review of the episodes, the device initially detected vt however the device was programmed to a vt-1 monitor zone between 160-193 bpm and the vt was occurring at 181 bpm.Within 2 minutes the patients ventricular rhythm had degenerated into either a low amplitude vf or agonal rhythm.The erratic ventricular activity was almost completely sensed appropriately but in very rare occurrences intermittent undersensing occurred due to variability in the egm amplitude.Several minutes later it was able to detect vf which caused the device to trigger a single round of atp.However before the device could charge, the degenerated rhythm caused the device to return to sinus.The ventricular activity was so low in amplitude that intermittent undersensing of the ventricular ectopic/agonal rhythm was occurring more frequently.All further episodes still present in the device memory and appeared to show the device in an atrial pace - bi-ventricular pace rhythm with apparent non-capture where it appears the patient was receiving cpr.
|