Model Number 397002-001 |
Device Problems
Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/29/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver has been returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the companion 2 driver exhibited a computer malfunction alarm during start up.
|
|
Manufacturer Narrative
|
The companion 2 driver was returned to syncardia for evaluation.The patient data file was reviewed and no computer malfunction alarms or any other relevant alarms were observed.Observation run and power cycle testing could not reproduce the customer-reported issue.The driver performed as intended, and there was no evidence of a device malfunction.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
|
|
Event Description
|
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the companion 2 driver exhibited a computer malfunction alarm during start up.
|
|