Information references the main component of the system and other applicable components are: product id: 3889-28, lot# va19u7w, product type: lead.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
A manufacturing representative (rep) reported that during a system replacement they were ran impedance at 2.6v and 360pw and they were getting over 4000ohms o all combinations, except a few case combinations.When they tested the lead with the test cable they were able to get good bellows and toe responses under 2.0v.They already took out the lead, rinsed it, wiped it down and reinserted, but obtained the same results except c-0, c-1 and c-2 were in range.They stated there was no bovie near the system, no x-ray in the room and no lights directed at the system.They also used an ins to test the responses.Additional troubleshooting had the rep use rate-14, pw-210us, set cycling at 3s on/3s off.No softstart, increase amplitude with the clinician programmer.It was noted that bellows and toe flick were present at nominal amplitude values.They tried another bipole pair to confirm programming options on the lead.They tested combinations and received the following responses: 0-3 3.3v, to, less bellows; 1-3 3.0v, toe, bellows are hard to see; 2-3 2.8v, toe, bellows; 0-1, 3.0v, little toes, bellows.The rep ran impedance again at 2.6v 360pw and still got greater than 4000ohms on all combinations except a few case pairs.There were no symptoms reported.Follow up with the rep noted that they replaced the lead and the patient felt stimulation appropriately, post operation, in the vagina, at 2.0 or less on all first 4 standard programs.The rep stated they would be returning the lead.The ins was indicated for urinary dysfunction/sacral nerve stimulation/gastrointestinal/pelvic floor.
|