• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS EXTENSION SET; SET, EXTENSION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS EXTENSION SET; SET, EXTENSION, INTRAVASCULAR Back to Search Results
Model Number 20128E
Device Problems Break (1069); Failure to Disconnect (2541)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/17/2016
Event Type  Injury  
Manufacturer Narrative
Although requested, the affected product has not been received.A follow up report will be submitted with investigation results should the devices be received for evaluation.
 
Event Description
The customer reported that the tubing broke off inside the central line catheter and the patient had to have their central line replaced.There is no report of lasting patient harm.
 
Manufacturer Narrative
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
 
Manufacturer Narrative
The customer¿s report that the set broke off inside the central line catheter was confirmed.Visual inspection showed the male luer tip was broken off, with stress marks visible at the break site.The broken off male luer tip was found lodged into the female port of the non-bd extension set.Functional testing was not performed due to the broken male luer tip and cracked spin collar.The root cause of the breakage is unknown.
 
Event Description
The affected port had been cleaned with alcohol prior to being connected.The tubing broke 16 hours later at the completion of a tpn infusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS EXTENSION SET
Type of Device
SET, EXTENSION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
ade ajibade
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key6153316
MDR Text Key61693936
Report Number9616066-2016-01749
Device Sequence Number1
Product Code FPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K801614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 11/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number20128E
Device Catalogue Number20128E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2016
Initial Date FDA Received12/07/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received12/29/2016
01/27/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CENTRAL LINE CATHETER, THERAPY DATE
Patient Outcome(s) Other; Required Intervention;
-
-