• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3357-40C
Device Problems Pacing Inadequately (1442); Under-Sensing (1661)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported a remote merlin transmission revealed a ventricular safety pacing recorded on an auto mode switching episode.Review of the episode revealed competitive atrial pacing resulting in intermittent undersensing.The patient will be monitored with routine follow-up.No further information is available.
 
Manufacturer Narrative
(b)(6).
 
Manufacturer Narrative
The reported field event of ventricular safety pacing recorded on an auto mode switching episode was not confirmed in the laboratory.The device was tested on the bench and no anomalies were found.
 
Event Description
New information received indicates that the device was explanted prophylactically.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6155668
MDR Text Key62206310
Report Number2938836-2016-15524
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2016
Device Model NumberCD3357-40C
Device Catalogue NumberCD3357-40C
Device Lot Number4691548
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/19/2016
Initial Date FDA Received12/08/2016
Supplement Dates Manufacturer Received10/19/2016
10/06/2017
Supplement Dates FDA Received09/22/2017
10/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age86 YR
-
-