• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Failure to Interrogate (1332); Communication or Transmission Problem (2896); Electromagnetic Compatibility Problem (2927); Human-Device Interface Problem (2949)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2016
Event Type  malfunction  
Event Description
Information was received from a manufacturer representative.The representative reported that they were unable to read the implantable neurostimulator (ins) intra-op and another ins was used due to the no telemetry.Post procedure the ins telemetry was good.It was discussed that the most likely cause was due to interference and it was ok to implant.The representative was going to try and use it in another procedure and noted that the ins was going to expire in one month.The ins was not implanted into the patient and there were no patient symptoms reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key6155700
MDR Text Key61883109
Report Number3004209178-2016-25433
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00613994913654
UDI-Public00613994913654
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2017
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Date Manufacturer Received12/07/2016
Date Device Manufactured07/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-