MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE PRO EVEROLIMUS ELUTING RX CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE PRO EVEROLIMUS ELUTING RX CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1017250-18
Device Problem Peeled/Delaminated (1454)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/15/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that the procedure was to treat a lesion in the mid left anterior descending artery.A 2.5x18mm xience pro rx stent delivery system (sds) was advanced toward the target lesion, the system was inflated and the stent was deployed.The stent system was withdrawn and an unusual polyethylene material was collected.The physician suspected that the coating on the catheter shaft peeled off while withdrawing the stent system.Reportedly the peeled off coating came out with the stent catheter.There was no resistance noted during advancement or withdrawal with the rotating hemostatic valve.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Internal file number - (b)(4): correction: device status changed from returning to not returned.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.Based on the information provided and without the product to examine, a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE PRO EVEROLIMUS ELUTING RX CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6156640
MDR Text Key62214850
Report Number2024168-2016-08727
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P070015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/20/2018
Device Catalogue Number1017250-18
Device Lot Number6031441
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2016
Initial Date FDA Received12/08/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-