Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/15/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The steerable guide catheter was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
|
|
Event Description
|
This is filed to report the loose connection of the stopcock to the steerable guiding catheter (sgc), which if used has the potential to cause or contribute to patient injury.It was reported that during preparation of the steerable guiding catheter (sgc), an attempt was made to connect a stopcock to the flush port of the sgc, but the stopcock continued to unscrew and would not stay on the device.Two additional stopcocks were attempted, but the same issue occurred.The sgc was not used and replaced.A new device was used without issue.There was no patient involvement with the first sgc and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
(b)(4).Evaluation summary: the device was returned with a stopcock and investigated.The reported loose/intermittent connection was unable to be confirmed.The returned device analysis confirmed that the steerable guide catheter (sgc) flush port functioned as expected as the returned stopcock was able to be securely tightened onto the flush port.A review of the lot history record revealed no manufacturing nonconformities.Additionally, a review of the complaint history identified no other incidents reported from this lot.All available information was investigated and the reported loose/intermittent connection is possibly related to variation in the non-abbott stopcocks used at the account; however, this cannot be definitively determined.Based on the information reviewed, there is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
|
|
Search Alerts/Recalls
|
|