• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0101
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Patient Involvement (2645)
Event Date 11/15/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The steerable guide catheter was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
This is filed to report the loose connection of the stopcock to the steerable guiding catheter (sgc), which if used has the potential to cause or contribute to patient injury.It was reported that during preparation of the steerable guiding catheter (sgc), an attempt was made to connect a stopcock to the flush port of the sgc, but the stopcock continued to unscrew and would not stay on the device.Two additional stopcocks were attempted, but the same issue occurred.The sgc was not used and replaced.A new device was used without issue.There was no patient involvement with the first sgc and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned with a stopcock and investigated.The reported loose/intermittent connection was unable to be confirmed.The returned device analysis confirmed that the steerable guide catheter (sgc) flush port functioned as expected as the returned stopcock was able to be securely tightened onto the flush port.A review of the lot history record revealed no manufacturing nonconformities.Additionally, a review of the complaint history identified no other incidents reported from this lot.All available information was investigated and the reported loose/intermittent connection is possibly related to variation in the non-abbott stopcocks used at the account; however, this cannot be definitively determined.Based on the information reviewed, there is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key6156978
MDR Text Key62245998
Report Number2024168-2016-08740
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2017
Device Catalogue NumberSGC0101
Device Lot Number60701U244
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/02/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-