MAUDE Adverse Event Report: EOC HEERLEN DISTRIBUTION CTR ATTAIN STABILITY; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EOC HEERLEN DISTRIBUTION CTR ATTAIN STABILITY; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 20066
Device Problems Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Vomiting (2144); Cardiac Tamponade (2226); Cardiac Perforation (2513); Pericardial Effusion (3271)
Event Date 09/20/2016
Event Type  malfunction  
Event Description
It was reported that during the implant procedure, patient experienced a pericardial effusion and cardiac tamponade due to a perforation.It was not known at which point the effusion occurred and what product was related to the effusion.It was noted the right atrial (ra) lead and right ventricular (rv) lead were placed and the physician inserted the balloon catheter into the coronary sinus, however, the x-ray did not show a brighter balloon silhouette as expected when the balloon was inflated.The physician performed the venogram and the fluid flushed away in the coronary sinus, as if the balloon was not inflated.The physician removed the balloon catheter and inflated the balloon outside the patient, however, it would not deflate after disconnecting the syringe and the valve.The balloon catheter was discarded and the physician chose to insert a different balloon catheter; it was added the physician may have repositioned the balloon while the balloon was already inflated within the coronary sinus.The venogram was performed and there were no signs of a perforation, dissection or contrast fluid within the pericardium at this point.The physician first chose to implant the left ventricular (lv) lead in a thinner antero-lateral vein, however, it was then decided to put the lv lead in the normal sized, postero-lateral vein and the lv lead was fixated.During slitting of the delivery catheter, the lv lead dislodged and slipped back into the coronary sinus and the lv lead implant attempt had to be restarted.In the meantime, both the right ventricular (rv) lead and right atrial (ra) lead fell off and needed to be repositioned.The second attempt to cannulate the coronary sinus with a new delivery catheter was more difficult and it was decided to implant an active fixation lv lead.It was noted at some point during the implant of the second lv lead, the patient status became critical for the first time; the patient's blood pressure was low, their skin color became whiter and the patient vomited.The physician then began to fixate the leads and implanted the cardiac resynchronization therapy pacemaker (crt-p).It was at the same time, another physician confirmed a pericardial effusion via ultrasound and the physician began pericardiocentesis to remove the blood from the pericardium.Since the device was not able to utilize telemetry c until after the pocket was closed, the leads were connected and the incision was seemed.It was then noticed the ra lead and rv lead were connected to the wrong ports in the crt-p and the pocket had to be re-opened.During the re-opening of the pocket, the ra and rv lead fell off again and needed to be repositioned; the lv lead then fell out of the coronary sinus.The patient status then became more critical and cardiopulmonary resuscitation (cpr) was performed, along with the use of a chest compression system for approximately thirty minutes.The physician then placed the existing ra lead into the right ventricle, and the rv lead and lv lead were explanted.It was decided to insert pin plugs into the ra port and lv port of the crt-p and utilize the chest compression system with medication until the patient was stable.The crt-p was programmed to vvi mode and the rv lead and crt-p remain in use.It was noted there was no evidence as to what could have caused the pericardial effusion and all of the attempted products were suspected to be related to the effusion.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN STABILITY
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
EOC HEERLEN DISTRIBUTION CTR
earl bakkenstraat 10
heerlen 6422 PJ
NL  6422 PJ
Manufacturer (Section G)
EOC HEERLEN DISTRIBUTION CTR
earl bakkenstraat 10
heerlen 6422 PJ
NL   6422 PJ
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6163385
MDR Text Key62202989
Report Number3002807561-2016-00019
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/01/2017
Device Model Number20066
Device Catalogue Number20066
Device Lot NumberYY0041722K
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/20/2016
Initial Date FDA Received12/09/2016
Supplement Dates Manufacturer ReceivedNot provided
09/20/2016
Supplement Dates FDA Received01/05/2017
09/26/2017
Date Device Manufactured09/09/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5076-52 LEAD, 4076-58, C2TR01 DEVICE
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age87 YR
-
-