• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 595000-001
Device Problems Device Alarm System (1012); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/28/2016
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the freedom driver exhibited a fault alarm, it continued to perform its life-sustaining functions.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited a fault alarm while supporting a patient.The customer also reported that the patient was switched to a back-up freedom driver without any adverse patient impact.
 
Manufacturer Narrative
The freedom driver was returned to syncardia for evaluation.The driver passed all pressure test requirements, which included cardiac output, rap (right atrial pressure), aop (aortic pressure), pap (pulmonary arterial pressure) and lap (left atrial pressure) performance metrics associated with nominal normotensive and hypertensive settings.Review of the electronic data revealed a fault code of "bottom pressure too high" which is indicative of a potentially failing u22 pressure sensor on the main pcba (printed circuit board assembly) and is consistent with the alarm that the customer experienced.Further investigation of the main pcba confirmed that the root cause for the reported fault alarm was a degraded pressure sensor (u22) on the main pcba.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 3603 follow-up report 1.
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited a fault alarm while supporting a patient.The customer also reported that the patient was switched to a back-up freedom driver without any adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6165954
MDR Text Key62452717
Report Number3003761017-2016-00385
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 11/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2016
Initial Date FDA Received12/12/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age64 YR
-
-