• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. UNIFY CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. UNIFY CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3235-40Q
Device Problem Defective Alarm (1014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/15/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
The device was explanted due to an inability to trigger the vibration alert during testing.
 
Manufacturer Narrative
Evaluation summary: interrogation of the device revealed the device was above eri when received.A longevity calculation was performed and found the battery depletion was normal based on the device usage.The device functionality was bench tested and no anomaly was detected.Based on this information, there was no evidence of device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6166486
MDR Text Key62129770
Report Number2938836-2016-16259
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2013
Device Model NumberCD3235-40Q
Device Catalogue NumberCD3235-40Q
Device Lot Number3535230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/31/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/26/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-