ACCLARENT, INC. RELIEVA SPIN SINUPLASTY SYSTEM; INSTRUMENT, ENT MANUAL SURGICAL Back to Search Results
Catalog Number RS0616F
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 09/02/2015
Event Type  malfunction  
Manufacturer Narrative
The age of the patient is not known.The user facility returned one (1) relieva spin balloon sinuplasty system with a relieva luma sentry¿ sinus illumination system, and an f-70 sinus guide catheter tip.The following observations were noted: the balloon catheter was wrinkled, and the tip was collapsed and detached from the shaft.Upon further inspection under the microscope, it was observed that the inner shaft of the sinus balloon broke off near the middle and a tear was noted on the proximal of the balloon.The shaft was also inspected under the microscope and the broken off edge appeared jagged which was likely caused by excessive force during use.Due to the condition of the returned device, testing of the balloon was not able to be replicated in the testing environment.Based on analysis findings, the reported complaint was confirmed.Product analysis was also conducted on the returned guide wire and guide catheter; however the observations and results of these are unrelated to the alleged complaint.Visual inspection of the guide wire found no anomalies.The guide wire was connected to the xenon light source to verify the light output.The light output was transmitted and the measurement was within specifications.The returned guide catheter was checked under the microscope and delamination on the blue tip was observed.A review of the lot history record for the subject device did not note any anomalies.
 
Event Description
Acclarent was informed that during a primary balloon sinuplasty procedure, the balloon catheter popped, detached and got stuck in the right maxillary sinus.The balloon catheter was replaced and the balloon popped again.Additional information was obtained and found that the balloon was completely detached and fell into the maxillary sinus.The balloon was completely retrieved using the takahashi forceps, no further surgery was performed.The patient was reported as doing well following the procedure; there were no complications noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIEVA SPIN SINUPLASTY SYSTEM
Type of Device
INSTRUMENT, ENT MANUAL SURGICAL
Manufacturer (Section D)
ACCLARENT, INC.
33 technology drive
irvine CA
Manufacturer Contact
joaquin kurz
33 technology drive
irvine, CA 92618
9497899383
MDR Report Key6168064
MDR Text Key62377622
Report Number3005172759-2016-00021
Device Sequence Number1
Product Code LRC
Combination Product (y/n)N
PMA/PMN Number
K111875
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 09/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/01/2017
Device Catalogue NumberRS0616F
Device Lot Number140401B-CM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/02/2015
Initial Date FDA Received12/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2014
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-