ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/28/2016
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of monitor sn (b)(4) has been completed.The reported problem (loose belt connection) was confirmed.As received, the monitor failed incoming testing.Upon evaluation, the belt receptacle was detached from the monitor case, damaging the internal pulse wire.The cause of the incoming test failure is the damaged belt receptacle.Root cause investigation is ongoing under an existing internal corrective action.No adverse event resulted from the defective monitor.
 
Event Description
A us distributor contacted zoll to report that the belt connection on a patient's monitor was loose.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
katelynn mains
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key6174723
MDR Text Key62694182
Report Number3008642652-2016-09151
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 12/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2016
Initial Date Manufacturer Received 12/07/2016
Initial Date FDA Received12/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-