Catalog Number 210804000 |
Device Problem
Break (1069)
|
Patient Problems
No Information (3190); No Code Available (3191)
|
Event Date 12/13/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
While drilling through cement, drill bit broke of in the patient.The piece was retrieved but, surgery was extended about 70 minutes.
|
|
Manufacturer Narrative
|
This complaint remains under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.
|
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Manufacturer Narrative
|
Examination of the returned instrument confirmed the complaint; the fluted tip broke off.The root cause is attributed to wear out.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the investigation, the need for corrective action is not indicated.Complaints will be monitored under post market surveillance sep 419.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|