• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARIBEX NOMAD PRO; EXTRAORAL SOURCE X-RAY SYSYTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARIBEX NOMAD PRO; EXTRAORAL SOURCE X-RAY SYSYTEM Back to Search Results
Model Number 0.850.0009
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 08/09/2016
Event Type  malfunction  
Manufacturer Narrative
Upon receipt of the complaint, it was determined that this was not a reportable event.Upon further review, it was determined this complaint is a reportable event.The investigation of the device has determined that a thermal event has occurred; however, further analysis is needed to verify root cause of the failure.A follow-up report will be submitted upon evaluation completion.
 
Event Description
It was reported that a battery is burnt and has a hole in it.
 
Manufacturer Narrative
An additional evaluation of the returned unit has been completed.Upon visual inspection it appears that a thermal event occurred.It is not possible to determine the exact sequence of events that led to the thermal event.However, there is evidence that a short occured in the lower battery pack between the cells and pcb.There was a blackening on the top of the lower battery pack and its printed circuit board (pcb).Cells 2 and 3 appear to have been involved in the thermal event, likely by supplying energy into a fault in the board.Additionally there was a slight melting of the handset plastic enclosure.This concludes the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO
Type of Device
EXTRAORAL SOURCE X-RAY SYSYTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key6177402
MDR Text Key62536893
Report Number1017522-2016-00044
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 08/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0009
Device Catalogue Number0.850.0009
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/09/2016
Initial Date FDA Received12/15/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-