• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE- IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE- IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1009K
Device Problem No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/15/2016
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume infusor experienced a no flow event during patient infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Device manufacture date: january 8, 2016 ¿ january 11, 2016.The actual device was not available; however, a photograph of the sample was provided for evaluation.For no flow problem, a functional flow rate test must be performed on the physical complaint sample in order to verify the flow rate accuracy of the device.Therefore, the reported no flow problem could not be objectively confirmed via the photos.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE- IRVINE
irvine CA
Manufacturer (Section G)
BAXTER HEALTHCARE- IRVINE
17511 armstrong ave
bldg 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6179481
MDR Text Key62535744
Report Number1416980-2016-18417
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C1009K
Device Lot Number16A006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/23/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-