The actual device was evaluated by the manufacturer for physical evaluation.An external, visual inspection of the returned cycler exterior showed no signs of any physical damage.A simulated treatment was completed without any failures or problems occurring.The cycler weighed fill volume values were within tolerance.The complaint symptoms could not be reproduced during testing.The valve actuation test passed.The system air leak test passed.The patient sensor calibration check passed.The load cell verification passed.The teach pumps test passed.The internal, visual inspection of the cycler encountered no discrepancies.Additionally, an investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the device history record review confirmed the labeling, material, and process controls were within specification.
|