|
Model Number H7493915010310 |
Device Problems
Fracture (1260); Difficult to Remove (1528)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/17/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.(b)(4).
|
|
Event Description
|
It was reported that catheter removal difficulties were encountered.The 90% stenosed target lesion was located in the severely tortuous and calcified internal carotid artery.After positioning a filter wire in the distal lesion, pre-percutaneous transluminal angioplasty was performed.A 10.0-31 carotid wallstent¿ was advanced to treat the lesion.However, during deployment when the stent had several centimeters more to be deployed completely, slight resistance was met.When the physician attempted to retract the stent into the sheath, the stent could not be retracted due to strong resistance.The device was then attempted to be retracted including the stent into the non-bsc guiding catheter but when the stent was pulled, the stent would not move and only the proximal delivery wire was pulled.The physician tried to push in the pulled delivery wire, then the outer sheath was moved and the stent was retracted.After that, the stent delivery system (sds) was advanced through the lesion together with the filter wire, and was then retracted into the guide catheter.Although the guide catheter got disengaged from the carotid artery, the sds crossed the lesion successfully and it was retracted into the guide catheter in the aorta.The procedure was then aborted.No patient complications were reported and the patient's status was good.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: an examination of the returned device found a break in the shaft.The break was located at the guidewire monorail exit on the distal outer.An examination of the break site found that the outer was stretched at the site of the break.This type of damage is consistent with excessive force being applied to the delivery system during device use.The distal end of the device including the stent was not returned for analysis.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
|
Event Description
|
It was reported that catheter removal difficulties were encountered.The 90% stenosed target lesion was located in the severely tortuous and calcified internal carotid artery.After positioning a filter wire in the distal lesion, pre-percutaneous transluminal angioplasty was performed.A 10.0-31 carotid wallstent¿ was advanced to treat the lesion.However, during deployment when the stent had several centimeters more to be deployed completely, slight resistance was met.When the physician attempted to retract the stent into the sheath, the stent could not be retracted due to strong resistance.The device was then attempted to be retracted including the stent into the non-bsc guiding catheter but when the stent was pulled, the stent would not move and only the proximal delivery wire was pulled.The physician tried to push in the pulled delivery wire, then the outer sheath was moved and the stent was retracted.After that, the stent delivery system (sds) was advanced through the lesion together with the filter wire, and was then retracted into the guide catheter.Although the guide catheter got disengaged from the carotid artery, the sds crossed the lesion successfully and it was retracted into the guide catheter in the aorta.The procedure was then aborted.No patient complications were reported and the patient's status was good.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was further reported that the shaft broke while the device was still inside the patient and the device was not intact when removed from the patient.
|
|
Search Alerts/Recalls
|
|
|