MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 QUICKSET 1PC FLEX DRILL BIT 35; HIP INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 QUICKSET 1PC FLEX DRILL BIT 35; HIP INSTRUMENT/TRIAL Back to Search Results
Catalog Number 227435500
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 12/14/2016
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Event Description
Drill bit broke inside pelvis and could not be retrieved.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Manufacturer Narrative
(b)(4).Examination of the returned drill confirmed the complaint; the majority of the fluted tip broke off.Review of the provided intra-op x-ray confirmed the broken piece is in the patient.A complaint database search has identified a trend for this failure within the (b)(4) quickset drill family.Previous investigations have determined that use error is the likely root cause.As a result of trending health hazard evaluation, (b)(4) was conducted.The investigation did not establish the need for corrective action.No evidence was found of product or design error as a contributing factor.Complaints will be monitored under sep (b)(4) post market surveillance.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET 1PC FLEX DRILL BIT 35
Type of Device
HIP INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6186105
MDR Text Key62754619
Report Number1818910-2016-33647
Device Sequence Number1
Product Code HTW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number227435500
Device Lot NumberPG243835
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/12/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/02/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age78 YR
-
-