Catalog Number GSX0030A
Device Problem Leak/Splash (1354)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 11/22/2016
Event Type  Injury  
Manufacturer Narrative
A review of the manufacturing records for the device verified that the lot met all pre-release specifications.The device was discarded, so no engineering investigation could be performed.(b)(4).
 
Event Description
It was reported the physician was implanting a gore® cardioform septal occluder in a defect that was balloon sized to 15-16mm.The device was locked and fully released.On echo the physician noted a residual leak and decided to remove the device.He advanced a 11f long arrow sheath and snare to remove the occluder.After the removal, echo demonstrated the atrial septum was modified due to the removal of the occluder.The septum was sized to 21mm in diameter with the balloon.The physician attempted to close the defect with a 24mm aso.The case was unsuccessful.The patient was taken to surgery for repair of the defect on (b)(6) 2016.The patient was doing well following surgery and was discharged from the hospital on (b)(6) 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® CARDIOFORM SEPTAL OCCLUDER
Type of Device
OCCLUDER, TRANSCATHETER SEPTAL
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
Manufacturer (Section G)
KENDRICK PEAK MPD B/P
4250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
ashley marostica
1500 n. 4th street
flagstaff, AZ 
9285263030
MDR Report Key6187059
MDR Text Key62794490
Report Number2017233-2016-00938
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2018
Device Catalogue NumberGSX0030A
Device Lot Number14819193
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age12 YR
Patient Weight28
-
-