Model Number 595000-001
Device Problem Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2016
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the freedom driver exhibited a noise, it continued to perform its life-sustaining functions.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited a noise while supporting a patient.The customer also reported that the patient was switched to a back-up freedom driver without any reported adverse patient impact.
 
Manufacturer Narrative
The freedom driver was returned to syncardia for evaluation.The customer-reported noise was not confirmed or reproduced during this investigation.The driver passed all test sections and pressure performance metrics associated with normotensive and hypertensive settings.Additionally, an extended observation run was performed, during which the driver functioned as intended with no anomalies, unintended alarms or noises.There was no evidence of a device malfunction.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Event Description
The customer, a syncardia authorized distributor, reported that the freedom driver exhibited a noise while supporting a patient.The customer also reported that the patient was switched to a back-up freedom driver without any reported adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6190652
MDR Text Key63202964
Report Number3003761017-2016-00404
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date 12/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/12/2016
Initial Date FDA Received12/20/2016
Supplement Dates Manufacturer Received10/19/2017
Supplement Dates FDA Received10/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age61 YR
-
-