Catalog Number MA-D20-G1
Device Problems Mechanical Problem (1384); Difficult To Position (1467)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during an unspecified surgical procedure, it was observed that the burr was not sticking in the angle attachment device.It was reported that there was an approximately five minute delay in the procedure, as an unspecified spare device was available for use and the procedure was successfully completed.It was reported that the device was successfully tested during pre-surgery set up.There was patient involvement reported.There were no patient or user injuries reported.It was reported there was no medical intervention or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was not confirmed.An assessment was performed and the device met all manufacture's specifications.Therefore, an assignable root cause was not determined.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINIMAL ACCESS ANGLE DRIVER, G1
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key6190939
MDR Text Key62910672
Report Number1045834-2016-13482
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133604
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMA-D20-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received01/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
BURR DEVICES
-
-