• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, 30CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, 30CM; SCS EXTENSION Back to Search Results
Model Number 3383
Device Problem Low impedance (2285)
Patient Problems Therapeutic Effects, Unexpected (2099); Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient¿s medical history and is unable to form an opinion as to the relevancy of the patient¿s history to the event reported.Sjm defers to the patient¿s physician regarding medical history.(b)(4).
 
Event Description
The patient had two extensions from the same lot.Device 2 of 2.Reference mfr report #: 1627487-2016-06514 it was reported the patient experienced stimulation turning off by itself.An impedance check revealed low impedance on multiple contacts.Troubleshooting was attempted but could not provide resolution.As a result, the ipg and extensions were explanted and replaced on (b)(6) 2016.The issue was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, 30CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key6191480
MDR Text Key62907611
Report Number1627487-2016-06516
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2013
Device Model Number3383
Device Lot Number3547100
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/09/2016
Initial Date FDA Received12/20/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL: 3228, SCS LEAD
Patient Outcome(s) Other;
-
-