Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr: returned product consisted of an epic self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination showed that the shaft was damaged approximately 10cm from the tip.There was also some buckling approximately 2cm from the tip.The stent was inadvertently deployed approximately 6mm from the marker band.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
Reportable based on device analysis completed on 07-dec-2016.It was reported that the stent failed to deploy.Vascular access was obtained via the femoral artery.The target lesion was located in the mildly tortuous and non-calcified superficial femoral artery (sfa).An 8x80x120 epic¿ stent was advanced to treat the lesion; however, the stent did not deploy.The physician tried to deploy the stent by rotating the thumb wheel both ways and by pull back method, but the stent never came out its sheath and the thumb wheel stopped rotating.The procedure was completed with a non-bsc stent and no patient complications were reported.However, returned device analysis revealed a partially deployed stent.
|