Catalog Number D2C4711K
Device Problem No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2016
Event Type  malfunction  
Manufacturer Narrative
Manufacturing date -- may 24, 2016 ¿ may 26, 2016.The device was returned and an evaluation is complete.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection was performed with no issues noted.A functional flow rate test was performed with no issues noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a small volume folfusor had an under infusion issue.The pump was filled with 5fu and nacl and after the infusion time was elapsed, 40ml remained in the pump.The total fill volume was 120ml.This was identified during use.There was no patient injury or medical intervention associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
irvine CA
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue, bldg 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6199699
MDR Text Key63133437
Report Number1416980-2016-18666
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 12/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Other
Device Catalogue NumberD2C4711K
Device Lot Number16E029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/29/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
5FU, NACL
Patient Age73 YR
-
-