MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Fluid/Blood Leak (1250)
Patient Problems ST Segment Elevation (2059); No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2016
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryoablation procedure a saline leak was noticed from the sheath valve and a wet area beneath it on the sterile cover was noted.The case was completed with cryo.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the sheath was returned and visually inspected and functionally tested.Visual inspection of the sheath showed the shaft was intact with no apparent issues.Air aspiration was reproduced when a catheter was introduced through the sheath.Dissection showed the hemostatic valve was leaking; the valve was torn.The reported ¿air bubble¿ issue was confirmed through testing; the sheath failed the inspection due to a torn hemostatic valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire, quebec H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire, quebec H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6200780
MDR Text Key63199705
Report Number3002648230-2016-00535
Device Sequence Number1
Product Code DRA
UDI-Device Identifier00643169688872
UDI-Public00643169688872
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number85595
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-