SMITHS MEDICAL ASD, INC. PORTEX® PEDIATRIC DISPOSABLE ANESTHESIA BREATHING CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101678D-NL |
Device Problems
Disconnection (1171); Loose or Intermittent Connection (1371)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/29/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Smiths medical has received the sample device.A full evaluation is anticipated, but not yet begun as the device is currently in transit to the investigation site.Smiths medical will file a follow-up report detailing the results of the evaluation once it is completed.(b)(4).
|
|
Event Description
|
It was reported that at the beginning of the procedure while performing induction, the breathing circuit disconnected between the elbow and the y piece of the mask during holding of the mask.It was reported that a leak test was performed prior to use.There were no adverse health outcomes reported.
|
|
Manufacturer Narrative
|
One device was returned for evaluation.Visual inspection found that the elbow and y-connector were not disconnected.Functional testing involved a force test and found the device met with iso specifications and passed the force test.A review of the testing and inspection documents was performed and found to be adequate.A review of the manufacturing process was performed with a similar part (same process procedures, process controls, product tests, line equipment, process flow, and components) and found no disconnections between elbow and the y-connector.Based on the evidence, the root cause was unable to be confirmed as the device met manufacturing specification.
|
|