• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3367-40QC
Device Problem Defective Alarm (1014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the device was explanted and replaced because the patient could not feel the vibratory notifier and the device was under advisory.No early depletion was noted.No further information was provided.
 
Event Description
New information states that patient is in stable condition and recovering.
 
Manufacturer Narrative
The reported field event of a patient notifier anomaly was not confirmed in the laboratory.The device was tested on the bench and no anomalies were found.
 
Manufacturer Narrative
Correction: - this supplemental report is to correct the previously submitted explant date.The explant date should have been (b)(6)2016.Correction: - this supplemental report is to correct the previously submitted report to include pma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6200969
MDR Text Key63175675
Report Number2938836-2016-21744
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup,Followup,Followup
Report Date 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2016
Device Model NumberCD3367-40QC
Device Lot Number4727453
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/01/2016
Initial Date FDA Received12/22/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
09/26/2018
Supplement Dates FDA Received01/10/2017
02/28/2017
10/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-