AV-TEMECULA-CT SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM Back to Search Results
Catalog Number SE-05-080-120-6F
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Aneurysm (1708); Thrombosis (2100)
Event Date 09/29/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.The reported patient effects of aneurysm and thrombosis, as listed in the supera, instructions for use are known patient effects associated with peripheral stenting procedures.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that on (b)(6) 2016 the 5.0x80 supera stent was implanted in the popliteal artery with a drug (paclitaxel) coated balloon.A control angiogram on (b)(6) 2016 showed in-stent thrombus, which was treated with thrombolysis.On (b)(6) 2016 an aneurysm in the popliteal was noted and the supera stent was floating in the vessel.The physician thought the paclitaxel from the drug coated balloon used during the index stent procedure may have contributed to the aneurysm.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Type of Device
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6201447
MDR Text Key63175696
Report Number2024168-2016-09217
Device Sequence Number1
Product Code NIP
UDI-Device Identifier08975370022346
UDI-Public(01)08975370022346(17)180228(10)6031961
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P120020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 12/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2018
Device Catalogue NumberSE-05-080-120-6F
Device Lot Number6031961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2016
Initial Date FDA Received12/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
3.5 MICROGRAM PACLITAXEL
Patient Outcome(s) Hospitalization; Required Intervention;
-
-