• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number ASKU
Device Problems Improper or Incorrect Procedure or Method (2017); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2016
Event Type  malfunction  
Manufacturer Narrative
This is a report of a use error where the patient did not make a secure connection between the patient line and the transfer set resulting in a disconnection.The disconnection between the patient line and the transfer set is the cause for the alarm.Use errors and proper user instructions are addressed in ¿the homechoice and homechoice pro systems patient at-home guide¿, which is shipped with every homechoice device.The guide instructs the user to check all disposable set connections for a secure fit before beginning therapy.Also, it provides step-by-step instructions for properly connecting the patient line to the transfer set.A review of the label for the product family will be conducted.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a system error 2240 alarm (air in set/line) occurred on the homechoice device during peritoneal dialysis therapy.This event occurred during drain three of three.The patient was connected at the time of the alarm.The patient noticed that the patient line became disconnected from their transfer set due to a loose connection made by the patient.The patient stated that their bed was soaked.The technical service representative (tsr) had the patient close the clamps and cycle power on the homechoice.The patient stated that they already capped off the transfer set.The tsr assisted the patient in ending therapy and explained to the patient that they could finish therapy with manual bags.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NI
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE CORPORATION - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6202567
MDR Text Key63225446
Report Number1416980-2016-18710
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 12/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/30/2016
Initial Date FDA Received12/23/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
HOMECHOICE
Patient Age65 YR
-
-