• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE- IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE- IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1063KP
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(6) the device was manufactured july 14, 2016 ¿ july 16, 2016.The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photograph revealed an indentation mark on the tube.The cause of the indentation mark could not be determined.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of a large volume infusor ¿became punctured¿ during patient infusion, leading to a leak.This occurred approximately one hour before the end of the expected therapy time.The device was filled with piperacillin / tazobactam in 0.9% sodium chloride solution.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE- IRVINE
irvine CA
Manufacturer (Section G)
BAXTER HEALTHCARE- IRVINE
17511 armstrong ave
bldg 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6203650
MDR Text Key63260433
Report Number1416980-2016-18713
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K071222
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 12/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number2C1063KP
Device Lot Number16G011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/29/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
PIPERACILLIN / TAZOBACTAM; SODIUM CHLORIDE 0.9%
-
-