MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) FORTIFY DR, U1.6 SJ4 US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) FORTIFY DR, U1.6 SJ4 US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2231-40Q
Device Problems Premature Discharge of Battery (1057); Difficult to Interrogate (1331)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/07/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a remote transmission could not be completed.The patient was brought into clinic, and the device could not be interrogated.It was noted that the device exhibited premature battery depletion.The device was successfully explanted and replaced.The patient was stable before, during, and after the procedure.
 
Manufacturer Narrative
No conclusion code available.Final analysis found the reported premature battery depletion was confirmed in the laboratory.Based on all available parameter and usage information, device longevity was found to be below the expected limits.With an external power supply attached, the device functioned normally.No high current was detected during testing and the power consumption was normal.The original battery was returned to the vendor for further evaluation and analysis could not conclusively determine a root cause for the premature battery depletion.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.In further analysis of the battery, lithium clusters were observed, but at the time of analysis the clusters did not appear to be in a location or size that would be sufficient to cause an internal short of the battery.As a result, the cause of the premature battery depletion could not be conclusively determined.However, from these analyses, in the absence of other root causes, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY DR, U1.6 SJ4 US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6205156
MDR Text Key63305738
Report Number2938836-2016-22154
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 06/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2012
Device Model NumberCD2231-40Q
Device Lot Number3199713
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/03/2016
Initial Date FDA Received12/27/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received03/20/2017
06/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0115-2017
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
-
-