Model Number MMT-523LNAS
Device Problems Break (1069); Positioning Problem (3009)
Patient Problem Hyperglycemia (1905)
Event Date 12/02/2016
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.(b)(4).
 
Event Description
Customer reported via phone call that insulin pump was damaged as it was worn in bra.Customer reported that reservoir compartment was broken and reservoir did not stay in place.Customer's blood glucose was 415 mg/dl.Customer would contact healthcare professional for instruction on taking manual injection.
 
Manufacturer Narrative
The insulin pump was received with a completely broken reservoir tube lip and unable to perform basic occlusion test and occlusion test due to broken reservoir tube lip.However, the insulin pump had normal operating currents and passed unexpected restart error test, self-test, rewind test, displacement test, prime test and lcd excessive no delivery test.The insulin pump was received with minor scratched window, cracked lcd window, cracked case at display window corners, cracked battery tube threads and broken belt clip slot.The insulin pump was missing the reservoir tube o-ring and the end cap sticker.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6209523
MDR Text Key63419134
Report Number2032227-2016-54236
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 01/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-523LNAS
Device Catalogue NumberMMT-523LNAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/02/2016
Initial Date FDA Received12/28/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/20/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age54 YR
Patient Weight64
-
-