The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device was determined to meet functional and electrical performance specification requirements per rite testing.A short simulated therapy was successfully performed.Visual inspection confirmed that the power switch was pushed into the machine.The broken power switch was replaced to resolve the reported issue.Should additional relevant information become available, a supplemental report will be submitted.
|