MAUDE Adverse Event Report: COVIDIEN KANGAROO JOEY,PUMP W/CLMP,REFB; EPUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO JOEY,PUMP W/CLMP,REFB; EPUMP Back to Search Results
Model Number 383401
Device Problems No Display/Image (1183); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 12/28/2016.An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2016 that a customer had an issue with an epump.The customer states that the screen was blank and would not work.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed for the reported condition of a blank screen.The unit was triaged and the reported issue was confirmed.Liquid ingress caused the pcba to corrode, and is the result of customer misuse.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEY,PUMP W/CLMP,REFB
Type of Device
EPUMP
Manufacturer (Section D)
COVIDIEN
building # 10, 789 puxing road
shanghai
CH 
Manufacturer (Section G)
COVIDIEN
building # 10, 789 puxing road
shanghai
CH  
Manufacturer Contact
edward almeida
15 hampshire st
mansfield, MA 02048
5084524151
MDR Report Key6210489
MDR Text Key63711520
Report Number3006451981-2016-00741
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number383401
Device Catalogue Number383401
Device Lot NumberF1273262
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/09/2016
Initial Date FDA Received12/28/2016
Supplement Dates Manufacturer Received08/09/2016
Supplement Dates FDA Received10/25/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-