ST. JUDE MEDICAL, INC. QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3371-40Q
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
During a routine follow up, it was not possible to interrogate the device and an ecg displayed a spontaneous rhythm of approximately 45 bpm.The patient reported feeling a slow rhythm for a couple of months and did not feel a vibratory alert.The patient was hospitalized and the device was removed without complications due to suspected premature battery depletion.The patient is in good condition.The device was retained by the hospital but will be sent back to sjm.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.In further analysis of the battery, lithium clusters were observed, but at the time of analysis the clusters did not appear to be in a location or size that would be sufficient to cause an internal short of the battery.As a result, the cause of the premature battery depletion could not be conclusively determined.However, from these analyses, in the absence of other root causes, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6213337
MDR Text Key63534198
Report Number2938836-2016-15377
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2015
Device Model NumberCD3371-40Q
Device Lot Number4152948
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2016
Initial Date FDA Received12/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0117-2017
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-