• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS RINGLOC MALLORY HEAD HIP SYSTEM ACETABULAR SHELL 48MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS RINGLOC MALLORY HEAD HIP SYSTEM ACETABULAR SHELL 48MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Metal Shedding Debris (1804)
Patient Problem No Information (3190)
Event Date 12/01/2016
Event Type  Injury  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.(b)(4).
 
Event Description
During an initial total hip arthroplasty, the surgeon noted metal particles on the acetabular shell rim after implantation.It was reported the particles appeared to be from the porous coating of the shell.The area was cleaned and all visible particles were removed.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed.Device history record (dhr) was reviewed and no discrepancies were found.Review of complaint history determined that no further action is required as no trends were identified.Root cause was unable to be determined.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RINGLOC MALLORY HEAD HIP SYSTEM ACETABULAR SHELL 48MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6214170
MDR Text Key63586938
Report Number0001825034-2016-05384
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK030055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number13-104148
Device Lot Number915640
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2016
Initial Date FDA Received12/29/2016
Supplement Dates Manufacturer Received11/17/2017
Supplement Dates FDA Received11/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/16/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-