Model Number P3700E000059
Device Problem Self-Activation or Keying (1557)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/13/2016
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the right hand side rail control board needed to be replaced.Per the hill-rom service manual the affinity three birthing bed and affinity four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance (pm) along with a quarterly battery check and testing for (b)(6).Pm and testing not only meet (b)(6) requirements but will help ensure a long, operative life for the bed.Pm will minimize downtime due to excessive wear.Check the switches in the side rails to ensure they are functioning correctly.Also check for intermittent operation.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2016.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the right side rail control board to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the head section will self run up.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED FRAME
Type of Device
BIRTHING BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key6215366
MDR Text Key63829239
Report Number1824206-2016-00552
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 12/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP3700E000059
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received12/13/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-