TRIAGE TOTAL 5 CONTROL L1; MULTI-ANALYTE CONTROLS Back to Search Results
Model Number 88753
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/10/2016
Event Type  malfunction  
Manufacturer Narrative
Mdr is being filed due to field action ref-2027969-10/12/16-004-c.Customer's complaint was replicated with in-house testing of the code chip of retain lot c3233a.Code chip ranges did not match ev card ranges.The manufacturing records for the control lot were reviewed and the lot met release specifications.A capa, (b)(4), was initiated to address this issue.Date of this report selected as 9/29/2016 to reflect the date in which field corrective action 2027969-10/12/16-004-c was initiated.
 
Event Description
The support consultant attempted to run the total 5 control level 1 at 2 stdev with a d-dimer result outside of the expected range low at 388 ng/ml.(2 stdev: 390-584 ng/ml).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIAGE TOTAL 5 CONTROL L1
Type of Device
MULTI-ANALYTE CONTROLS
Manufacturer Contact
ya-ling king
9975 summers ridge road
san diego, CA 92121
8588052084
MDR Report Key6220289
MDR Text Key63813298
Report Number2027969-2017-00002
Device Sequence Number1
Product Code JJY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072892
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date 09/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number88753
Device Lot NumberC3233A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/10/2016
Initial Date FDA Received01/03/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction Number2027969-10/12/16-004-C
Patient Sequence Number1
Treatment
TRIAGE D-DIMER PANEL, PN 98100, LOT# W62095B; TRIAGE METERPRO DOMESTIC KIT, PN 55070, SN (B)(4)
-
-