• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 397002-001
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 12/16/2016
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because it did not prevent the companion 2 driver from performing its life-sustaining functions.The companion 2 driver has been returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver compressors cycled on and off while the driver was connected to wall air and supporting a patient.The customer also reported that the patient was subsequently switched to the backup companion 2 driver.There was no reported adverse patient impact.
 
Manufacturer Narrative
The companion 2 driver was returned to syncardia for evaluation.During investigation testing, a malfunction of the manual pressure regulator to maintain the required pressure set point value was identified as the root cause of the customer-reported compressor cycling while the driver was connected to wall air.The issue of compressor cycling because of a malfunction of the manual pressure regulator is being investigated in a capa (corrective or preventive action).The capa will document the investigation including, but not limited to, potential root cause and corrective actions associated with compressor cycling because of a malfunction of the manual pressure regulator.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver compressors cycled on and off while the driver was connected to wall air and supporting a patient.The customer also reported that the patient was subsequently switched to the backup companion 2 driver.There was no reported adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6221957
MDR Text Key64182454
Report Number3003761017-2016-00414
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 12/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/16/2016
Initial Date FDA Received01/04/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age54 YR
-
-