Model Number NOT APPLICABLE
Device Problems Fluid/Blood Leak (1250); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2016
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.A batch record review of kit lot e206 was conducted.There were no nonconformances associated with this lot.The lot met release requirements.Trends were reviewed for complaint categories, tubing leak and alarm #57: return pump (#3) error and no trend was detected for these categories.The assessment is based on information available at the time of the investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.(b)(4).
 
Event Description
Customer called to report an alarm #57: return pump (#3) error at 360 ml of whole blood processed (wbp).Customer stated tubing was stiff when putting it on the pump.Alarm could be reset after reloading pump segments.Customer noticed blood leak close to the filter.Treatment was aborted.Patient was stable and did not need any extra fluids.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS
hampton NJ 08827
Manufacturer (Section G)
THERAKOS, INC.
10 north high street, suite 30
west chester PA 19380
Manufacturer Contact
megan vernak
53 frontage road
po box 9001
hampton, NJ 08827
MDR Report Key6223015
MDR Text Key64175119
Report Number2523595-2017-00003
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date08/01/2018
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXECP
Device Lot NumberE206
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/06/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
Patient Weight75
-
-