Catalog Number LF2-R090
Device Problems Material Puncture/Hole (1504); Material Separation (1562); Use of Device Problem (1670); Device Handling Problem (3265)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 12/15/2016
Event Type  malfunction  
Manufacturer Narrative
-the concerned device "lacrifast" is not distributed in us under this device name, but is identical to the device "lacriflow lacrimal stent" distributed in us under 510(k) # k120886.- the actual device used was returned and investigated: there was a hole, from where the bougie was thought to penetrate out, in the sidewall of the tube (colored in blue) where a stainless steel ring was mounted, and the ring was dismounted and missing.The ring's whereabouts was not confirmed during the procedure in this (b)(6), and accordingly, the missing stainless steel ring might have been remaining in the patient body.- as a possible cause of the penetration of the bougie and dropping off the ring during the use, we speculate as follows: when the doctor tried to insert and advance the tube with the bougie inserted forcibly in the occluded lacrimal duct, excessive mechanical force was loaded on the ring and the distal tip of the tube where the ring was mounted, and further pushing the bougie resulted in dismounting the ring and penetration of the bougie with the ring out of the tube.
 
Event Description
This device (lacrifast) was employed to treat the epiphora due to lacrimal duct obstruction.The doctor inserted one of the lacrimal duct tube (colored in blue) into the lacrimal duct.While advancing the tube, the doctor felt that the bougie penetrated out of the tube and retrieved it out of the patient's lacrimal duct.He used another lacrifast and completed the operation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LACRIFAST
Type of Device
LACRIMAL STENT
Manufacturer (Section D)
KANEKA CORPORATION
2-3-18
nakanoshima, kita-ku
osaka, osaka 530-8 288
JA  530-8288
Manufacturer Contact
kazuhiko inoue
2-3-18
nakanoshima,kita-ku
osaka, osaka 530-8-288
JA   530-8288
4613072
MDR Report Key6224507
MDR Text Key63918147
Report Number9614654-2017-00001
Device Sequence Number1
Product Code OKS
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2019
Device Catalogue NumberLF2-R090
Device Lot NumberKP076298
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/16/2016
Initial Date FDA Received01/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-