• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. QUADRA ASSURA DDQP+; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. QUADRA ASSURA DDQP+; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3267-40Q
Device Problem Vibration (1674)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/06/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
During a follow-up, the patient was not able to feel the vibratory alert, even when the alert worked correctly.The device was reprogrammed to increase the duration of the vibratory alert with no resolution.The device was explanted and replaced and the patient was stable.
 
Manufacturer Narrative
The device was tested in the lab and no anomalies were found.The patient notifier was also tested and revealed no anomaly.A longevity calculation was performed and found the battery depletion was normal based on the device usage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA DDQP+
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6225104
MDR Text Key63948298
Report Number2938836-2017-00302
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2015
Device Model NumberCD3267-40Q
Device Lot Number4201136
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2016
Initial Date FDA Received01/05/2017
Supplement Dates Manufacturer Received09/28/2017
Supplement Dates FDA Received10/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age83 YR
-
-